INX-500 mA X-ray machine

Technichal Specification

Parameter Specified value
Mains Voltage 415VAC
Frequency 50Hz
Line impedance   1.5Ω max
Length of power cable 3.0 meter
Power output 24.0 kw
Output frequency Two Pulse
Kvp range 40 – 125 kvp in 1KV/Step
Kvp accuracy ≤ ± 5%
mAs range 1 – 320 mAs
mAs accuracy ≤ ± 10%
Overload protection Programmed
Mains isolation Power supply cable shall be plugged where both poles (L & N) are isolated.
Earthing Earthing N TO G <5V
x-ray tube Rotating anode
Focal spot nominal  1.0mm & 2.0mm
Inherent tube filtration 0.9mm
Target angel 16˚
Collimator Manual
Exposure release switch 2 steps
Type Floor to ceiling FIX
Mobility safety lock  Brake locking option on column
Environmental condition 
For transport & storage:
Temperature range 
  -15˚C to +45˚C  
Humidity (RH) Upto 90% (no condensation)
Atmospheric pressure. 50kpa to 105kpa
Operating :  
Temperature 10˚c to 40˚c
Humidity (RH) Upto 85% (non condensing)
Conforms to standards IS:7620   PART-1
AERB approved

Product Description

Vertex, a 500 mA X-ray machine by Innovation Meditech Pvt. Ltd. With 415VAC mains voltage, 24.0 kW power output, and 1.0mm/2.0mm focal spot, it ensures precise diagnostics. AERB approved and compliant with IS:7620 PART-1, it guarantees reliable imaging in diverse medical settings. Proudly manufactured by INNOVATION MEDITECH (P) LTD., a leading supplier of X-ray machines in India.

Additional Information

Delivery Time: 15 Days

Scroll to Top